Radon Detector Corentium Pro – High-precision radon monitor for the professional
Corentium Pro, preferred by radon professionals and home inspectors. AARST-NRPP certified for the North American market and beyond. Four highly accurate radon chambers work in parallel to achieve maximum reliability, performance, accuracy and speed.
This is the professional radon detector for home inspectors and any company that wants to offer professional radon measurements as a service to protect people in homes, students in schools or employees of companies from possible health damage caused by radon.
Details of the Radion Detector Corentium Pro
Functions
Air quality measures: radon, temperature, humidity
Corentium Pro samples indoor air through a passive diffusion chamber, using alpha spectrometry to precisely calculate the radon level. Radon is detected using silicon photodiodes to both count and measure the energy of alpha particles resulting from the decay chain of radon gas. The instrument is calibrated to reference instruments in accredited laboratories and is AARST-NRPP certified.
Environmental sensors
- Temperature: 0.2°C resolution, ± 1°C accuracy
- Humidity: 0.5% RH resolution, ± 4.5% accuracy
- Barometric pressure: 0.002 kPa resolution, ± 1 kPa accuracy
Specifications
- Radon sampling: passive diffusion chamber
- Detection method: alpha spectrometry
- Detector: 4 silicon photodiodes in 4 distinct radon chambers
- Diffusion time constant: 25 min
- Measurement range: 0 - 50 000 Bq/m3
- Sensitivity: ~100 cph at 1000 Bq/m3
- Sampling rate: 1 hour
- Operation environment
- • 4°C to 40°C
- • 5% RH to 85% RH non-condensing
- • 50 kPa to 110 kPa
Memory
- Internal memory stores 5 years of data
- Capacity: ~ 1900 days of measurement
- 5 data sets of one year length
- 177 data sets of 1-week length
- 325 data sets of 2-days length
- Memory type: non-volatile flash memory
Long battery life
- Approximately 1.5 year of continuous monitoring
- Comes with replaceable AA batteries
Data & Accuracy
- Data accessible via Android or iOS app
- Free reporting and analysis software for PC
The standard deviation from the exact radon level is
- less than 7% + 5 Bq/m3 after 24 hours (US: 7% + 0.12 pCi/L) and
- less than 5% + 2 Bq/m3 after 7 days (US: 5% + 0.05 pCi/L)